日付: | 09/26/2019 |
カテゴリー: | MultiQuant Software , PeakView Software |
For research use only. Not for use in diagnostic procedures.
Issue Description
After upgrading PeakView™ software to version 2.2 or MultiQuant™ software to version 3.0.3, the following error message was generated: "The procedure entry point nextafterf could not be located in the dynamic link library MSVCR120_CLR0400.dll."
Resolution
When PeakView or MultiQuant software are upgraded, the .NET Framework may also be upgraded to .NET 4.7.2, which may cause the software to stop responding. The workaround is to remove .NET Framework 4.7.2 and then install the .NET 4.0 software provided with the Multiquant 3.0.3 and Peakview 2.2 software installation packages.
Posted: now