What Value Should Be Used for the XIC Width in My SWATH® Acquisition MicroApp Processing Settings?

One common error when using the SWATH® Acquisition MicroApp 2.0.1 for processing SWATH data is to relate the XIC width setting in the processing software with the mass accuracy of the instrument, and set the XIC width to a value that is too low. XIC extraction widths should be set to 50 to 75 ppm, regardless of the mass accuracy of the acquired data.

Are CloudConnect and SWATH® MicroApps within PeakView® Software Windows 10 Compatible?

The software package, PeakView® 2.2, CloudConnect 1.2 beta and SWATH® 2.0.1 software. is needed to perform the OneOmics workflow and/or SWATH protein data processing and is Windows 7 and Windows 10, 64-bit compatible.

Licensing Information for the SWATH Acquisition™ MicroApp in PeakView® Software

When viewing license information in the PeakView® software Help menu, the SWATH™ acquisition microapp licensing info shows in the plug-ins list, but not in the PeakView License Report screen.

Can the SWATH™ Acquisition Variable Window Calculator v 1.0 Be Used with All TripleTOF® Systems?

SWATH™ acquisition software and the variable window methods will work on the TripleTOF® 4600, 5600 and 6600 systems, but Analyst® TF 1.7 software must be installed.

12/15/2017    Analyst Software | SWATH Acquisition

How Is the SWATH® Acquisition Variable Window Calculator Excel Sheet Used?

This article includes step-by-step instructions on how to calculate variable windows for your SWATH® acquisition experiment and how to generate the input and output list.

How Are SWATH™ Acquisition Files Recalibrated?

The procedure for recalibrating a SWATH™ acquisition file is outlined here.