How to Interface MultiQuant™ Software with a LIMS System via a LIMS Report Template

Oftentimes MultiQuant™ software results table data needs to be exported and then interfaced with a LIMS system. This article explains how to indirectly transfer data from results tables created in MultiQuant software to a LIMS system via a report template. The report template creates an output file in CSV format that can then be interfaced with a LIMS system.

11/01/2021    Clinical | Mass Spectrometers

Obtaining a Smaller Volume Divert Valve for a 4500, 5500 and 6500 LCMS Systems

To obtain a smaller volume divert valve for a 4500, 5500 or 6500 LCMS system, a new part must be ordered from Valco, part number of C52-2006I.

05/13/2020    Mass Spectrometers

How to Determine if the Acquisition PC Has a Bad Ethernet Card?

This article summarizes some issues that might contribute or confirm a bad network card.

04/16/2020    Mass Spectrometers

What Is the Settling Time in an Analyst® Software Acquisition Method?

In Analyst® software, the settling time is defined as the time that a mass spectrometer pauses after finishing a cycle until it begins the next cycle. The instrument electronics require a settling time in order to function successfully.

11/09/2019    Mass Spectrometers

Longer Power Cycle May Help To Resume Instrument Connectivity after Power Failure

Even after resetting the mass spectrometer (by following the power-cycling procedure), the instrument would not come back online after a power failure. On some occasions, the mass spectrometer must be completely powered down for at least 1 min in order for the circuits to discharge and fully release any faulty information.

10/01/2019    Mass Spectrometers

Where Can the Software for StatusScope® Remote Monitoring Service Be Accessed?

To obtain access to the software download for StatusScope® Remote Monitoring service, users who are under warranty or who have purchased the service can navigate to statusscope.absciex.com and select the link to the software download.

09/29/2019    Mass Spectrometers

How to adjust, save and export the table settings in an Analyst Quantitation Table

Instructions how to modify (choose and adjust the visible columns, change the significant figures and notation), save and export an Analyst Quantitation Table.

Ordering Foam Applicators for Cleaning a IonDrive™ QJet Ion Guide

The IonDrive™ QJet ion guide can be cleaned with an anti-static foam swab (part number 1017396 or 1017398).

10/18/2018    Mass Spectrometers

Using the 25 Micron ESI Electrode with the DuoSpray™ Source?

A 25 micron ESI electrode can be used with the IonDrive™ Turbo V source and the DuoSpray™ source.

Using Ammonium Fluoride (AMF) in an HPLC Solvent

Ammonium fluoride can be used in LC/MS applications as a mobile phase additive. It is recommended that only small concentrations as possible be used (around 0.2 mM).

07/03/2018    Mass Spectrometers

What Are the Reasons for Hybrid Electrodes Probe Blockage?

Common issue reported with hybrid electrodes probes.

05/15/2018    Academia Omics | Mass Spectrometers

Integrated Syringe Error Message Recieved while Using Manual Tune Mode in Analyst® Software

While running Analyst® software in Manual Tune mode during compound optimization, an error message is received: "The current sample cannot be run because the syringe pump has stalled. The syringe plunger may be stalled in the syringe barrel." To resolve this issue, move the syringe holder back and clear error. If this doesn't work, replace the syringe, or if still needed, reset the MS.

12/21/2017    Mass Spectrometers

What Files Must Be Backed Up before an Analyst Software Reinstall?

Back up the API Instrument project folder in the Analyst Data folder prior to any reinstall or upgrade of the Analyst software. To upgrade an older version of Analyst, remove and back up the Analyst Data folder and any remaining short cuts on your workstation. Reboot the PC and install the new version of the software.

Injection volume must be a positive value

If the Regional Settings in an Acquisition Computer with Analyst® Software is not set to US English , this can course the Error message: "Batch not submitted-Injection volume must be a positive value" during submitting a Batch.

How Do You Set Up for Manual Tuning Using Tee Infusion?

Before setting up your manual tuning method for source/gas optimization, your MS instrument must be set up for continuous infusion with high LC flow.

10/13/2017    Mass Spectrometers

A summary of the findings from an investigation into the identity of background Ions found in -ve mode electrospray mass spectrometry with suggestions to help minimize their impact on analysis.

10/02/2017    Mass Spectrometers

What is my Source Exhaust and how does it work?

Description of the Source Exhaust components and common requirements to maintain optimal performance

09/12/2017    Mass Spectrometers

Power cycling the PC, HPLC, and MS to solve communication problems

A power cycle can be performed on LCMS components to solve communication problems.

03/20/2017    Mass Spectrometers

How to wire SCIEX AUXIO cable for use with a Waters Acquity HPLC System

Diagram of how to wire AUXIO connections for the Acquity HPLC including the Stop Flow on instrument error

06/02/2016    Mass Spectrometers

Setup of Diverter Valve

How to setup a diverter valve, including plumbing and Analyst software setup.

Tuning 4500,5500 and 6500 using Instrument Optimization

Walk through of instrument optimization

02/05/2016    Mass Spectrometers

Venting and Pumping Down the 5500 Series Mass Spectrometer

How to remove the vacuum and pump the machine back down afterwards

02/05/2016    Mass Spectrometers

Submitting and Starting Sample Batch in Analyst® Software

How to start acquiring data using Analyst software

Create Acquisition Method with Agilent HPLC in Analyst® Software

Walk Through of Analyst acquisition method (.dam) creation when a Agilent HPLC is employed.

Create Acquisition Method with Shimadzu HPLC in Analyst® Software

Walk through of creating an Analyst acquisition method (.dam) when a Shimadzu HPLC is employed.