ナレッジベース資料(FAQ)


ナレッジベース資料は、よくある質問に対する回答を見つけるためのFAQです。

最近の記事


How to Interface MultiQuant™ Software with a LIMS System via a LIMS Report Template

Oftentimes MultiQuant™ software results table data needs to be exported and then interfaced with a LIMS system. This article explains how to indirectly transfer data from results tables created in MultiQuant software to a LIMS system via a report template. The report template creates an output file in CSV format that can then be interfaced with a LIMS system.

11/01/2021    Clinical | Mass Spectrometers

Can Multiple Users Log In to Windows while Analyst® MD Software Is Acquiring Data?

If a user starts a batch using Analyst® MD Software and then signs out of their Windows account while Analyst Software is acquiring data, the software will close and suspend the run. It is not recommended to sign out of a user's Windows account if Analyst MD Software is acquiring data.

12/02/2020    Clinical | Analyst Software

Is the MPX™ Driver Software System Compatible with Analyst® MD software?

The MPX™ driver software system is not supported with Analyst® MD software. Compatible LC systems are listed in the installation guide for Analyst MD software.

03/20/2020    Clinical | Forensics | Analyst Software

List of CE Capillaries with Pre-burned Detection Windows

CE capillaries with pre-burned detection windows are available for purchase; the dimensions and coating information for these products are listed here.

Changing the Number of Decimals Displayed in Analyst® Software Calibration and Metric Plots

This is a set of instructions for changing the number of decimals displayed in Analyst® software quantitation metric plots and calibration curves. The default number of decimals leaves the significance so low that you can not reproduce the results shown in the results table. A warning is given to let customers know that Regedit is a dangerous utility.

How to adjust, save and export the table settings in an Analyst Quantitation Table

Instructions how to modify (choose and adjust the visible columns, change the significant figures and notation), save and export an Analyst Quantitation Table.

Removing Analytes from a MultiQuant Software Method (.qmethod)

To remove an unneeded analyte's MRM transitions from a MultiQuant software method, the quantitation method must be opened and the Delete Selected Rows function must be used.

Starting an Analyst Software Data File (.wiff) with a Space

Can an Analyst® software data file name start with a space? The answer is yes. It is not a good practice to start the file name (.wiff) with a space, but Analyst software does allow the user to do that by using the format of "Subfolder name\ Datafile name" in the data column of the batch.

05/25/2018    Clinical | Analyst Software

How to Generate MRM Ratios in Analyst Software

This is a known issue in the ProteinPilot™ software 5.0.2 and was resolved with a patch.

How Do You Setup a Solenoid or Solvent Selection Valve on a Shimadzu LC?

This article describes how to setup a FCV-11 AL valve for a Shimadzu LC system.

11/22/2017    Clinical

What Is the Function of the Collision Gas (CAD) on QTRAP® System Instruments?

The various functions of the CAD gas are described for the QTRAP® System instruments and the different scan types.

How Can the Data from Two Experiments Acquired within One Period Be Separated in Analyst Software?

Data from two or more experiments acquired within one period can be separated by opening a data file in Explore mode and double clicking the green arrow.

06/06/2017    Clinical | Software

How Are the Method Parameters Chosen for Autocalibration Samples in the X500 Series?

Method parameters for autocalibrations submitted in SCIEX OS batches are automatically chosen from the sample after the autocalibration.

Generating Excel Reports in Chemoview 2.0.2 Fails With Message Programmatic Access to Visual Basic Project is not trusted

When creating excel reports using Chemoview 2.0.2 the report creation will fail and the message Programmatic Access to Visual Basic Project is not trusted will be displayed in the report preview.

06/02/2016    Clinical | ChemoView Software

How Do I Import and Export ChemoView Settings

ChemoView settings can be Exported so that they can be backed up or transferred between instruments or computers using the Import function.

12/22/2015    Clinical | ChemoView Software