I am running Analyst 1.5.1 with or without June 2010 hotfixes and I am having failed acquisitions with my batches I generate using a HPLC utilising an AAO Analyst Access Objects driver.|質量分析計・LC-MS/MSシステムのSCIEX


日付: 09/04/2016
カテゴリー: Analyst Software

0 投票
   印刷する    記事を評価する:

For research use only. Not for use in diagnostic procedures​
Answer

There are two patches available one for June 2010 hotfixes and one without June 2010 hotfixes installed. Both patches correct an issue found in the Analyst® 1.5.1 software that prevents users from using the Quantitation option available within the Batch Editor when the selected quantitation method contains more than 150 transitions and control of any LC device is established through AAO (Analyst® Access Objects). In this scenario, attempting to acquire the batch results in failed acquisitions. Please submit request through http://sciex.com/request-support to request this patch.