Answer

Changing the options available for individual parameter settings in Analyst® software (e.g., changing the CAD gas settings to display just “Low – medium – high” and not numeric values) can be done by accessing the following menus:

  • Select Configure or Tune mode. Then select Tools > Settings > Parameter Settings.
  • Choose the scan mode.
  • Double click on the corresponding parameter. (For the above example, double click on CAD.)
  • Click on Default for Access type, which changes Access type to Simplified.
  • The only parameter values available for the Simplified type are Low – Medium – High.
  • Click OK to save.