Can I display the pressure trace of a chromatogram when I use an Agilent LC system with Analyst|質量分析計・LC-MS/MSシステムのSCIEX


日付: 06/02/2016
カテゴリー: Analyst Software

0 投票
   印刷する    記事を評価する:

For research use only. Not for use in diagnostic procedures.
Answer 

The pressure trace has to be activated in the hardware profile.

With deactivated profile, go to “Edit Profile”, click on the pump, then on “Setup Device”, and activate the checkbox “Enable Real-Time Monitoring”

When you acquire data with this hardware profile, the pump pressure can be displayed in Analyst Explore mode with right mouse click “Show ADC data” (in PeakView software: Show – UV/DAD/ADC channel)