Chromatograms Are Not Appearing in Reports Created with Reporter Template Suite|質量分析計・LC-MS/MSシステムのSCIEX


日付: 11/22/2021
カテゴリー: Analyst Software

0 投票
   印刷する    記事を評価する:

For research use only. Not for use in diagnostic procedures.


Issue Description

When using MultiQuant™ or SCIEX OS software to create a report with Reporter Template Suite software, chromatograms are not appearing in the report. Instead, the report shows the text "This image is not available" where the image should appear.

User-added image


Resolution

If the tag, Picture: Peak Review, is used, then the Label field must contain a value that matches the name of a column in the results table or it must be blank. Chromatograms will not appear in a report if the Peak Review picture tag contains a numerical value in the Label field (e.g.,12, an unknown Results Table column name.)

The Label field should be either empty or match a results table column name with an underscore character between words (e.g., Analyte_Peak_Area). 


User-added image