Why Is No Data Acquired in EAD Fragmentation Mode on a ZenoTOF 7600 System?|質量分析計・LC-MS/MSシステムのSCIEX


日付: 11/02/2021
カテゴリー: SCIEX OS

0 投票
   印刷する    記事を評価する:

For research use only. Not for use in diagnostic procedures.


Answer

If EAD fragmentation is used in an acquisition method on a ZenoTOF 7600 system, then the accumulation time must be at least three times the reaction time. If it is not, then no data will be acquired. To resolve the issue, increase the accumulation time in the method in SCIEX OS software.