Are PPGs Used for Tuning and Calibrating the QTRAP® or TripleQuad™ LC-MS/MS 7500 Systems?|質量分析計・LC-MS/MSシステムのSCIEX


日付: 12/04/2020
カテゴリー:

1 投票
   印刷する    記事を評価する:

For research use only. Not for use in diagnostic procedures.


Answer

No, a new calibrant, the MS Tuning Solution, was introduced for tuning and calibrating the QTRAP® or TripleQuad™ LC-MS/MS 7500 Systems. This new MS Tuning Solution is used to tune both positive and negative mode in QqQ and linear ion trap (LIT) modes. The MS Tuning Solution is used when running the automated tuning and calibration workflow in the SCIEX OS Software.